Performance Ver
ification
Table 38: DVM Voltage Accuracy (DC) (cont.)
Channel 4
1
Vertical S cale
Input Voltage
Offset Voltage
Low Limit Test Result High Limit
1
–5 –5
–5.117 –4.883
0.5 –2 –2 –2.052 –1.948
0.5 –1 –0.5 –1.0345 –0.9655
0.2 –0.5 –0.5 –0.5175 –0.4825
0.01 0.002 0 0.00042 0.00442
0.2 0.5 0.5 0.4825 0.5175
0.5 1 0.5 0.9655 1.0345
0.5 2 2 1.948 2.052
1
55
4.883 5.117
1
Channels 3 and 4 are only on four-channel oscilloscopes.
68 MDO3000 Series Specifications and Performance Verification