Commands Listed in Alphabetical Order
WFMInpre:NR_Pt
This command specifies the number of data points that are in the incoming
waveform record to be sent to the oscilloscope using the
CURVe command.
(See page 2-95, Waveform Transfer Command Group.)
Group
Waveform Transfer
Syntax
WFMInpre:NR_Pt <NR1>
WFMInpre:NR_Pt?
Related Commands
CURVe,
DATa,
DATa:STARt,
DATa:STOP,
SAVe:WAVEform,
SAVe:WAVEform:FILEFormat,
WFMOutpre:NR_Pt?
Arguments
<NR1> is the number of data points if WFMInpre:PT_Fmt is set to Y. It is the
number of min-max pairs if WFMInpre:PT_Fmt is set to ENV.
Examples
WFMINPRE:NR_PT 10000 specifies that 10000 data points will be expected.
WFMINPRE:NR_PT ? might return :WF MINPRE:NR_PT 10000 indicating that
there are 10000 data points in the expected incoming waveform record.
WFMInpre:PT_Fmt
This command specifies the acquisition format of the data points to be sent tothe
oscilloscope using the
CURVE command. This can be Y for YT format, or ENV
for envelope mode (min/max pairs). For YT format, each data value represents a
single waveform data point. For envelope format, each data point represents a
min/max pair, where the minimum value precedes the maximum value.
(See page 2-95, Waveform Transfer Command Group.)
Group
Waveform Transfer
2-950 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual