Commands Listed in Alphabetical Order
Frequency
Function Min Max Resolution
SINE
0.1Hz 50MHz 0.1Hz
SQUare
PULSe
ARBitrary
0.1Hz 25MHz 0.1Hz
LORENtz
GAUSsian
ERISe
EDECAy
NOISe
1
DC
1
HAVERSINe
0.1Hz 5MHz 0.1Hz
SINC (Sin
(x)/x)
0.1Hz 2MHz 0.1Hz
CARDIac
RAMP
0.1Hz 500kHz 0.1Hz
1
DC and NOISe waveforms do not use the frequency setting. Frequency settings that are set when the
AFG:FUNCtion is DC or NOISe are remembered and applied when the function is changed to other than DC
or NOISe.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:FREQuency <NR3>
AFG:FREQuency?
Arguments
Floating point number that represents the AFG frequency, in Hz.
Returns
Query response is returned in high precision NR3 format (up to 12 digits with
more than 4 trailing 0 digits after the decimal point is omitted).
Examples
AFG:FREQUENCY 100.0E3 sets the AFG frequency to 100 kHz.
AFG:FUNCtion
Sets (or queries) which AFG function to execute.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-131

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals