Commands Listed in Alphabetical Order
The following v
alues can be set for low level:
AFG:LOWLevel Setting Value
Load Impedance
FIFTY Ohm
Load Impedance
HIGHZ
Min Max Resolution Min Max Resolution
SINE
SQUare
PULSe
RAMP
NOISe
DC
1
HAVERSINe
CARDIac
ARBitrary
-2.5V 1.24V 1mV -5.0V 2.49V 2mV
SINC (Sin(x)/x)
-1.518V 1.248V 1mV -3.036V 2.496V 2mV
LORENtz
-1.233V 1.266V 1mV -2.466V 2.532V 2mV
GAUSsian
ERISe
EDECAy
-1.25V 1.24V 1mV -2.5V 2.5V 2mV
1
When the :AFG:FUNCtion is specified as DC, a change to the AFG:LOWLevel value results in a corresponding change to :AFG:OFFset. A corresponding
change to :AFG:AMPLitude is remembered but is not applied until the :AFG:FUNCtion is changed to a function other than DC.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:LOWLevel <NR3>
AFG:LOWLevel?
Arguments
Floating point number that represents the AFG low level value, in volts.
Examples
AFG:LOWLEVEL 1.0 sets the AFG low level value to 1.00 volts.
AFG:NOISEAdd:PERCent
Sets (or queries) the AFG additive noise level as a percentage. Minimum is 0.0%,
maximum is 100.0% and increment is 1.0%.
2-136 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual