Commands Listed in Alphabetical Order
The maximum noi
se percent is limited when AFG:AMPLitude is greater than
50% of its maximum setting value for the current function, in accordance with
the following formula:
Noise Percent Max = 100.0 * (Amplitudemax/Amplitude 1.0)
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:NOISEAdd:PERCent <NR3>
AFG:NOISEAdd:PERCent?
Arguments
Floating point number that represents the AFG additive noise level, as a
percentage.
Examples
AFG:NOISEADD:PERCENT 50 sets the AFG additive noise level to 50 percent.
AFG:NOISEAdd:STATE
Sets (or queries) the AFG additive noise state.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:NOISEAdd:STATE {0|1|OFF|ON}
AFG:NOISEAdd:STATE?
Arguments
1 orON turns on the AFG additive noise state.
0 or OFF turns it off.
Examples
AFG:NOISEADD:STATE ON turns on the additive noise state.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-137

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals