Commands Listed in Alphabetical Order
AFG:OFFSet
Sets (or queries) the AFG offset value, in volts. The offset values that can be
specified for each function and load impedance are in the table below. (To
specify the f
unction and load impedance, use the commands AFG:FUNCtion
and AFG:OUTPut:LOAd:IMPEDance.)
NOTE. The AFG:OUTPut:LOAd:IMPEDance command must be used before
using any of the amplitude/offset/highLevel/lowLevel/Preset commands, so that
saved setups are recalled with the proper load factor applied.
AFG:OFFSet Setting Value
Load Impedance
FIFTY Oh
m
Load Impedance
HIGHZ
Min Max Resolution Min Max Resolution
SINE
SQUare
PULSe
RAMP
NOISe
DC
SINC (
Sin(x)/x)
GAUSsian
LORENtz
ERIS
e
EDECAy
HAVERSINe
CARD
Iac
ARBitrary
-1.25V 1.25V 500uV -2.5V 2.5V 1mV
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:OFFSet <NR3>
AFG:OFFSet?
Arguments
Floating point number that represents the AFG offset, in volts.
2-138 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals