Commands Listed in Alphabetical Order
Group
Status and Erro
r
Syntax
*RST
Related Commands
FACtory,
RECAll:SETUp,
SAVe:SETUp,
*PSC,
*RCL,
*SAV
Arguments
None
Examples
*RST resets the oscilloscope settings to factory defaults.
*SAV (No Query Form)
Stores the state of the oscilloscope to a specified memory location. You can use
the *RCL command to restore the oscilloscope to this saved state at a later time.
Group
Save and Recall
Syntax
*SAV <N R1>
Related Commands
*RCL, RECAll:SETUp,
SAVe:SETUp
Arguments
<NR1> specifies a location in which to save the state of the oscilloscope. Location
values range from 1 through 10. Using an out-of-range location value causesan
execution error. Any settings that have been stored previously at this location
will be overwritten.
Examples
*SAV 2 saves the current oscilloscope state in memory location 2.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-593