Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:ARINC429A:ERRTYPE
This command specifies the error type when triggering on an ARINC429
bus signal. The trigger condition needs to be set to ERROR (using
TRIGger:A:B
US:B<x>:ARINC429A:CONDition). B<x> is the bus number.
Conditions
This command requires installation of a DPO4AERO application module for
MDO4000C models, or MDO3AERO for MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:ARINC429A:ERRTYPE {ANY|GAP|WORD |PARity}
TRIGger:A:BUS:B<x>:ARINC429A:ERRTYPE?
Arguments
ANY sets the error type to match any of the other available error types.
GAP sets the error type to match on gap violations (less than 4 bits idle time
between two packets on the bus).
WORD sets the error type to match on word errors (incorrect number of bits in a
word or bits that violate return to zero transmission).
PARity sets the error type to match on parity errors (parity value results in even
parity count for a word).
Examples
TRIGger:A:BUS:B1:ARINC429A:ERRTYPE ANY sets the error type to match
any of the other available error types.
TRIGger:A:BUS:B1:ARINC429A:ERRTYPE? might return PARity.
TRIGger:A:BUS:B<x>:ARINC429A:LABel:HIVALue
When the trigger condition is set to LABel, and the qualifier is set
to either INrange or OUTrange, this command specifies the upper
value of the range for a trigger on the label field. (Use the command
TRIGger:A:BUS:B<x>:ARINC429A:LABel:VALue to specify the lower value of
the range). The default is all X's (don't care). B<x> is the bus number.
Conditions
This command requires installation of a DPO4AERO application module for
MDO4000C models, or MDO3AERO for MDO3000 models.
Group
Trigger
2-778 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual