41
Model UT805: OPERATING MANUAL
Accuracy Specifications
Accuracy: (a% reading + b digits), guarantee for 1 year. Operating temperature: 23
0
C 5
0
C.
Relative humidity: not more than 75% RH. Temperature coefficient: 0.1 x (specified accuracy)/1
0
C.
A. DC Voltage
Remarks:
Overload Protection: 1000V DC or 750V AC
Input Impedance:
At 200mV and 2V ranges: more than 500M
At 20V, 200V and 1000V ranges: Around 10M
l
l
Range Measurement Scope Resolution Accuracy
200mV
2V
20V
200V
1000V
1 V~220.000mV
10 V~2.20000V
100 V~22.0000V
1mV~220.000V
10mV~1000.00V
1 V
10 V
100 V
1mV
10mV
In REL mode:
(0.006%+2)
(0.007%+2)
Note
UT805 multimeter is a precise digital multimeter when you turn on the UT805 allow a warm-up period of at least
30 minutes for the internal components to stabilize this ensures that the multimeter accord with the specification
l