▪ It is mandatory to ensure that the PC system
configuration is compatible with the PC system
requirements for the imaging and image viewer software.
▪ Since image quality may be deteriorated from lack of
resources, observe the requirement guideline specified in
the following tables.
▪ The PC components shall be approved by UL/CSA.
▪ The PC shall be grounded well protectively.
▪ The multiple portable socket-outlets shall not be placed
on the floor.
▪ In case the equipment is to be installed in an area with
an unstable electric power supply, it is strongly
recommended to use the AVR (automatic voltage
regulator) to keep the line voltage stable.
▪ The PC system provided with the PaX-i Plus / Insight
undergoes the rigorous test for software compatibility
before shipping. Therefore, any later changes to the
hardware and/or software may cause malfunction.