EasyManuals Logo
Home>ABB>Measuring Instruments>AO2000 Series

ABB AO2000 Series User Manual

ABB AO2000 Series
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
134 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
ZO23: Gas analyzer start-up
Gas analyzer start-up procedure, initial calibration at the installation site
1 Turn on the gas analyzer power supply.
After approx. 15 minutes, the sample cell has reached its operating
temperature. If required, the gas analyzer can be calibrated at the ref-
erence point (see step 3) and the end point (see step 5).
2 For setting the reference point (= electrical zero), feed ambient air wait
and until the measured value has stabilized (duration approx. 2 hrs.).
Meanwhile, purge the test gas valves and the gas supply line with oxy-
gen-free gas (e.g. with nitrogen from a loop feeder) or with sample gas
(flow rate 5 to 10 l/h).
3 Set reference point to 20.6 vol.% of O
2
.
4 Feed span gas and wait until the measured value has stabilized (dura-
tion max. 2 hrs.).
5 Set end-point value in accordance with the certificate of analysis of
6 The gas analyzer is ready for performing measurements; feed sample
gas.
NOTE
Information for test gases can be found in the section "ZO23: Preparation
for installation" (see page 67).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB AO2000 Series and is the answer not in the manual?

ABB AO2000 Series Specifications

General IconGeneral
BrandABB
ModelAO2000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals