208 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Configuring value entry
Configuring value entry
1 Select menu item Display.
2 Start configuration of the new display element with NEW.
3 Select the parameter Page.
The page overview is displayed.
4 Select the page on which the display element is to be displayed. Only
those pages can be selected with an assignment < 100 %, i.e. in which
there is at least one free position.
If the selected page is a system page, the position of the display ele-
ment is set by the system; it can only be changed using Swap display
(see page 205).
If the new page is a user page, the position must be configured.
5 Select the parameter Position.
The nine possible positions are graphically represented; free positions
are identified by the position number.
6 Select the position with the corresponding number key.
7 Select the parameter Type.
8 Select display type Value entry.
This creates a Constant function block; whose system-issued name,
'Value page-position', is displayed in the parameter Source. This name
cannot be changed here; it can only be changed by configuring the func-
tion block (see step 11).
9 Select the parameter Config entry and configure the other parame-
ters: entry range, decimal places, text and password level. The configu-
ration of reverse entry ranges (e.g. 100 to 0 ppm) is possible.
10 The description of the display element is entered in the parameter
Measuring point.
11 Select the function block created in step 8, enter the name and unit, and
link the function block to an application via its output (see Technical In-
formation "Function Blocks – Descriptions and Configuration" for de-
tailed description).
12 Switch to measuring mode.
The newly configured display element is now shown in the display. The
description of the display element is shown above the display. To the
right of the display, the name and unit of the function block are dis-
played that were entered in step 11.