AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 37
Installation location
Installation location requirements
The gas analyzer is only intended for installation indoors.
The technical data of the gas analyzer (see data sheet and section "Analyzer
module operating specifications" (see page 346)) are valid for an installation
location altitude up to 2000 meters above sea level.
The installation site must be stable enough to bear the weight of the gas
analyzer!
Short gas paths
Install the gas analyzer as close as possible to the sampling location.
Install the gas conditioning and calibration modules as close as possible to
the gas analyzer.
Adequate air circulation
Provide for adequate natural air circulation around the gas analyzer. Avoid
heat build-up.
When installing several system housings in a 19-inch housing, maintain a
minimum spacing of 1 height unit between housings.
The complete surface of the system housing is used to dissipate the heat
losses.
Protection from adverse conditions
Protect the gas analyzer from
• Cold,
• Exposure to heat from e.g. the sun, furnaces, boilers
• Temperature variations
• Strong air currents
• Accumulation of dust and ingress of dust
• Corrosive atmosphere
• Vibration.