EasyManuals Logo
Home>ABB>Measuring Instruments>AO2000 Series

ABB AO2000 Series User Manual

ABB AO2000 Series
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #215 background imageLoading...
Page #215 background image
AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 215
Air pressure effect
If the gas analyzer does not have an integrated pressure sensor for air pres-
sure correction, the air pressure value should be checked and reset (see
page 272) as required prior to calibrating the analyzer module:
• if the gas analyzer’s operating site altitude has changed since the last
calibration or
• if the air pressure effect (see page 346) on the measured value is too
high.
Waiting period following manual calibration
If the Output Current Response parameter is set to Hold, current output
is halted for a specific time to allow the measurement value to stabilize after
automatic calibration is ended.
This interval is:
Test gas → Sample gas purge time + 4 x T90 or
Test gas → Sample gas purge time + 1 x T90-1 + 3 x T90-2.
The waiting period is the same as that following automatic calibration (see
page 216).
Calibration data
Setting the calibration data is described in section "Calibration data for
manual calibration" (see page 226).
Analyzer module manual calibration
Manual calibration of an analyzer module is described in section "Analyzer
module manual calibration" (see page 265).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB AO2000 Series and is the answer not in the manual?

ABB AO2000 Series Specifications

General IconGeneral
BrandABB
ModelAO2000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals