EasyManuals Logo
Home>ABB>Measuring Instruments>AO2000 Series

ABB AO2000 Series User Manual

ABB AO2000 Series
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #262 background imageLoading...
Page #262 background image
262 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Analyzer with internal cross-sensitivity correction
During calculation possible electronic cross-sensitivity and/or carrier gas
corrections by other measurement components are disabled for the calcula-
tion. The following information should therefore be noted in particular:
All the sample components must always be calibrated in the following se-
quence for the zero-point calibration:
• First that sample component which is not corrected,
• Afterwards that sample component which is affected by the smallest
number of corrections,
• Up to that sample component which is affected by the largest number
of corrections.
Example
Sample components CO
2
, CH
4
, C
3
H
6
Cross-sensitivity correction CO
2
by CH
4
,
CO
2
by C
3
H
6
,
C
3
H
6
is not corrected.
Sequence for zero-point calibration 1. C
3
H
6
, 2. CH
4
, 3. CO
2
.
All the sample components must also always be calibrated in the end-point
calibration. Here, a corrected sample component may only be calibrated
using a test gas contain no components which cause cross-sensitivity, i.e.
which only consists of the sample component and an inert gas, e.g. N
2
.
Wait until the warm-up phase has ended
The analyzer module should only be calibrated after the warm-up phase (see
page 135).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB AO2000 Series and is the answer not in the manual?

ABB AO2000 Series Specifications

General IconGeneral
BrandABB
ModelAO2000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals