08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
9.4 Initial setup on the meter
There are two ways to customize the setup of the meter:
directly on the meter (see the following menu overview)
or via the data management system (see Appendix A).
The two methods differ according to the range of options
available.
The meter is delivered with preconfigured default set-
tings. The following settings can be configured on the
meter:
■ Date and time format
■ Date and time editing enabled
■ Units
■ Test types
■ Lot information editing enabled
■ Methods for verifying lot information
■ Enable barcode scanner
■ Normal, critical, and reportable ranges
■ Operator ID options
■ Patient ID options
■ Settings for glucose controls and QC Lockout
■ Beeper
If the meter is configured by a system administrator
using a data management system, the configuration
options on the meter may be disabled to avoid conflicts
in the settings. These options then appear grayed out
on the display.
The option Service Menu (see section 10.7) with the
feature “Delete Patient Information” is not available in
the preconfigured default setting. The Service Menu
can only be activated via DMS.