If wireless connectivity is activated, the (RF OFF)
icon is displayed as a button in the Power Up screen.
■ To temporarily disable the wireless network con-
nection, touch (RF OFF) in the Power Up
screen. The icon then changes to (RF ON).
■ To temporarily enable the wireless network con-
nection, touch (RF ON) in the Power Up
screen. The icon then changes to (RF OFF).
The RF ON/RF OFF button always displays the option you
currently have. The current communication status is dis-
played in the status bar (bottom line) of the display in all
menus and screens.
■ The icon is displayed, if wireless connectivity is
enabled. Signal strength is shown by 4 bars next to
the icon in the status bar. Signal strength can
range from “poor” to “excellent” .
■ The icon is displayed if the last attempt to com-
municate with the data management system was
successful and was terminated according to the
communication protocol.
■ The icon is displayed if the last attempt to
communicate with the data management system
was not successful or was unintentionally termi-
nated. If this icon persists, contact your system
administrator. Ignoring this information may lead to
a Download Lockout (if configured, see page 48).
The signal strength bars are only displayed when wire-
less communication is actually taking place. For example,
since wireless communication never takes place when a
test is being run on the meter, no status bars will be
displayed.