To unsecure the function generator for calibration
The function generator can use a calibration security code to prevent
unauthorized or accidental calibration. This procedure shows you how to
unsecure the function generator for calibration.
Shift 1 Turn on the menu.
Menu On/Off A: MOD MENU
< 2 Move across to the CAL MENU choice on this level.
F: CAL MENU
¿
3 Move down a level to the SECURED command.
1: SECURED
If the display shows UNSECURED, you do not need to perform this
procedure to execute a calibration.
3
Chapter 3 Front-Panel Menu Operation
To unsecure the function generator for calibration
47