Trigger Mode/Coupling 11
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 183
Note that Trigger Coupling is independent of Channel Coupling (see "To
specify channel coupling" on page 65).
To enable or disable trigger noise rejection
Noise Rej adds additional hysteresis to the trigger circuitry. By increasing
the trigger hysteresis band, you reduce the possibility of triggering on
noise. However, this also decreases the trigger sensitivity so that a slightly
larger signal is required to trigger the oscilloscope.
1 Press the [Mode/Coupling] key.
2 In the Trigger Mode and Coupling Menu, press the Noise Rej softkey to
enable or disable.
To enable or disable trigger HF Reject
HF Reject adds a 50 kHz low- pass filter in the trigger path to remove high
frequency components from the trigger waveform.
You can use HF Reject to remove high- frequency noise, such as AM or FM
broadcast stations or noise from fast system clocks, from the trigger path.
1 Press the [Mode/Coupling] key.
2 In the Trigger Mode and Coupling Menu, press the HF Reject softkey to
enable or disable.
Pass
Band
50 kHz
DC
0 dB
-3 dB

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals