260 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
17 Waveform Generator
The waveform generator output is always disabled when the instrument
is first turned on.
The waveform generator output is automatically disabled if excessive
voltage is applied to the Gen Out BNC.
2 In the Waveform Generator Menu, press the Waveform softkey and turn
the Entry knob to select the waveform type.
3 Depending on the selected waveform type, use the remaining softkeys
and the Entry knob to set the waveform's characteristics.
Waveform
Ty pe
Characteristics Frequency
Range
Max.
Amplitude
(High-Z)
1
Offset
(High-Z)
1
Arbitrary Use the Frequency/Frequency
Fine/Period/Period Fine, Amplitude/High-Level,
and Offset/Low-Level softkeys to set the arbitrary
waveform signal parameters.
Use the Edit Waveform softkey to define the
arbitrary waveform shape. See "To edit arbitrary
waveforms" on page 263.
100 mHz to
12 MHz
20 mVpp to
5Vpp
±2.50 V

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals