Agilent InfiniiVision 3000 X-Series Oscilloscopes
User's Guide
s1
15
Mask Testing
To create a mask from a "golden" waveform (Automask) 243
Mask Test Setup Options 245
Mask Statistics 248
To manually modify a mask file 249
Building a Mask File 252
One way to verify a waveform's compliance to a particular set of
parameters is to use mask testing. A mask defines a region of the
oscilloscope's display in which the waveform must remain in order to
comply with chosen parameters. Compliance to the mask is verified
point- by- point across the display. Mask test operates on displayed analog
channels; it does not operate on channels that are not displayed.
To enable mask test order Option LMT at time of oscilloscope purchase, or
order DSOX3MASK as a stand- alone item after oscilloscope purchase.
To create a mask from a "golden" waveform (Automask)
A golden waveform meets all chosen parameters, and it is the waveform to
which all others will be compared.
1 Configure the oscilloscope to display the golden waveform.
2 Press the [Analyze] key.
3 Press Features; then, select Mask Test.
4 Press Features again to enable mask testing.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals