Reference 22
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 333
Transient Withstand Capability
Environmental Conditions
Maximum input voltage at analog inputs
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
50
Ω input: 5 Vrms Input protection is enabled in 50 Ω mode and the 50 Ω load will
disconnect if greater than 5 Vrms is detected. However the inputs could still be
damaged, depending on the time constant of the signal. The 50
Ω input protection only
functions when the oscilloscope is powered on.
With 10073C 10:1 probe: CAT I 500 Vpk
With N2862A or N2863A 10:1 probe: 300 Vrms
Maximum input voltage at digital channels
±40 V peak CAT I; transient overvoltage 800 Vpk
Environment Indoor use only.
Ambient
temperature
Operating 0 °C to +55 °C; non-operating –40 °C to +71 °C
Humidity Operating: Up to 80% RH at or below +40 °C. Up to 45% RH up to +50 °C.
Non-operating: Up to 95% RH up to +40 °C. Up to 45% RH up to +50 °C.
Altitude Operating and non-operating to 4,000 m (13,123 ft)
Overvoltage
Category
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
Pollution Degree The InfiniiVision 2000/3000 X-Series oscilloscopes may be operated in
environments of Pollution Degree 2 (or Pollution Degree 1).