256 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
16 Digital Voltmeter
The DVM makes accurate RMS measurements when the signal frequency is
between 20 Hz and 100 kHz. When the signal frequency is outside this
range, "<BW Limit?" or ">BW Limit?" appears in the DVM display to
caution you about inaccurate RMS measurement results.
To use the digital voltmeter:
1 Press the [Analyze] key.
2 Press Features; then, select Digital Voltmeter.
3 Press Features again to enable the DVM measurements.
4 Press the Source softkey and turn the Entry knob to select the analog
channel on which digital voltmeter (DVM) measurements are made.
The selected channel does not have to be on (displaying a waveform) in
order for DVM measurements to be made.
5 Press the Mode softkey and turn the Entry knob to select the digital
voltmeter (DVM) mode:
AC RMS — displays the root- mean-square value of the acquired data,
with the DC component removed.
DC — displays the DC value of the acquired data.
DC RMS — displays the root- mean- square value of the acquired data.
Frequency — displays the frequency counter measurement.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals