EasyManuals Logo
Home>Agilent Technologies>Test Equipment>N9310A

Agilent Technologies N9310A User Manual

Agilent Technologies N9310A
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
Subsystem Command Reference 6
N9310A User’s Guide 109
Amplitude Subsystem
The amplitude subsystem controls the settings and
parameters associated with amplitude.
Continuous Wave Output Amplitude
:AMPLitude:CW <val> <unit>
:AMPLitude:CW?
This command sets the amplitude of CW output.
The query returns the current amplitude of CW output.
Range and
unit
127 to +13 dBm, (+20 dBm settable)
80 to +60 dBmV, (+67 dBmV settable)
20 to +120 dBμV, (+127 dBμV settable)
0.0001 to 1000 mV, (2238.8 mV settable)
0.1 to 1 000 000 μV, (2238800 μV settable)
Minimum
increment
0.1 dB (When using dBm/dBmV/dBμV)
0.1 mV (When using mV)
0.1 μV (Ampl 999.9 μV)
100 μV (Ampl 1000 μV)
*RST value 127.0 dBm
Returned
data format
<val> <unit>
Front panel
access
Amplitude

Table of Contents

Other manuals for Agilent Technologies N9310A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9310A and is the answer not in the manual?

Agilent Technologies N9310A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9310A
CategoryTest Equipment
LanguageEnglish

Related product manuals