EasyManuals Logo
Home>Agilent Technologies>Test Equipment>N9310A

Agilent Technologies N9310A User Manual

Agilent Technologies N9310A
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
48 N9310A User’s Guide
3Using Functions
Generating an LF Output
The signal generator allows you to generate an LF (Low
Frequency) signal with the following characters. Low
frequency signal is also usually called audio frequency.
Operation Example
Assume you are going to generate an LF signal with:
a frequency of 10 kHz
•an amplitude of 3 V
Refer to the following steps to generate the LF signal:
1 Pressing enters submenu of LF output function.
2 Pressing LF Out Freq > 10 > kHz
sets LF frequency to 10 kHz
3 Pressing LF Out Ampl > 3 > V
sets LF amplitude to 3 V
4 Pressing LF Out On/Off to On state enables LF output.
Character Range Default
Frequency 20 Hz to 80 kHz 1.0000 kHz, with 0.1 Hz resolution
Amplitude 0 to 3 V (peak) 500 mV, with 1 mV resolution
Output
Connection
LF OUT connector on the
front panel (BNC type, 50 W)
N/A
LF Out

Table of Contents

Other manuals for Agilent Technologies N9310A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9310A and is the answer not in the manual?

Agilent Technologies N9310A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9310A
CategoryTest Equipment
LanguageEnglish

Related product manuals