Subsystem Command Reference 6
N9310A User’s Guide 129
FM Deviation
:PM:DEViation <val> <unit>
:PM:DEViation?
This command sets the FM deviation.
Choices for the variables <val> depend on FM rate.
The query returns the current FM deviation.
FM Rate
:PM:RATE <val> <unit>
:PM:RATE?
This command sets the frequency of the internal FM source.
The query returns the current frequency of the internal FM
source.
Range 0 to 10 rad (300 Hz < FM rate < 10 kHz)
0 to 5 rad (10 kHz < FM rate <
20 kHz)
Valid unit rad
Minimum increment 0.001 rad
*RST value 0.000 rad
Returned data format <val> <unit>
Front panel access FM > FM Deviation
Range 300 Hz to 80 kHz
Minimum increment 0.1 Hz
*RST value 1.0000 kHz
Returned data format <val> <unit>
Front panel access FM > FM Rate