Subsystem Command Reference 6
N9310A User’s Guide 131
Pulse Period
:PULM:WIDTh <val> <unit>
:PULM:WIDTh?
This command sets the pulse period of the internally
generated pulse modulation signal.
The query returns the current pulse period of the internal
pulse source.
Pulse Width
:PULM:PERiod <val> <unit>
:PULM:PERiod?
This command sets the pulse width of the internally
generated pulse modulation signal.
The query returns the current pulse width of the internal
pulse source.
Range 200 ms to 2s
Valid unit s, ms, us
Minimum increment 1 us
*RST value 200 us
Returned data format <val> <unit>
Front panel access Pulse > Pulse Period
Range 100 us to 1s
Valid unit s, ms, us
Minimum increment 1 us
*RST value 100 us
Returned data format <val> <unit>
Front panel access Pulse > Pulse Width