72 N9310A User’s Guide
4Key Reference
Pulse Period Pressing this softkey modifies the pulse period of the
internal pulse signal. The active function area displays the
current pulse period value. The allowed value range is 200
μs to 2 s, with the minimum increment of 1 μs. The pulse
period applies only to the internal pulse signal. If an
external pulse source is selected, the Pulse Period softkey is
disabled.
Default value: 200 μs
Key sequence: > Pulse Period
Pulse Width Pressing this softkey modifies the pulse width for the
internal pulse signal. The active entry area displays the
current pulse width value. The allowed value range is 100 μs
to 1 s, with the minimum increment of 1 μs. If an external
pulse source is selected, the Pulse Width softkey is disabled.
Default value: 100 μs
Key sequence: Pulse > Pulse Width