Characteristics and Specifications 5
Electrical Specifications
U1231A/U1232A/U1233A User’s Guide 109
Capacitance specifications
Tab le 5- 3 Capacitance specifications
Range Resolution
Accuracy
Measuring rate
(at full scale)
U1231A U1232A U1233A
1000 nF 1 nF 1.9% + 2 1.9% + 2 1.9% + 2
4 times/second10 μF0.01 μF 1.9% + 2 1.9% + 2 1.9% + 2
100 μF0.1 μF 1.9% + 2 1.9% + 2 1.9% + 2
1000 μF1 μF 1.9% + 2 1.9% + 2 1.9% + 2 1 time/second
10 mF 0.01 mF 1.9% + 2 1.9% + 2 1.9% + 2 0.1 times/second
Notes for capacitance specifications:
1 Overload protection: 600 Vrms for short circuits with <0.3 A current.
2 The accuracy of for all ranges is specified based on a film capacitor or better, and after the Null function is used to
subtract the residual values (by opening the test leads).
3 The maximum display is 1200 counts.