7 Performance Tests and Calibration
Calibration Overview
This manual contains procedures for verification of the
instrument’s performance and adjustment (calibration).
Closed - case Electronic Calibration
The instruments features closed- case electronic calibration.
No internal mechanical adjustments are required. The
instrument calculates correction factors based upon the
input reference value you set. The new correction factors are
stored in non- volatile memory until the next calibration
adjustment is performed. Non - volatile EEPROM calibration
memory does not change when power has been off.
Agilent Technologies Calibration Services
When your instrument is due for calibration, contact your
local Agilent Service Center for a low- cost recalibration. This
product is supported on automated calibration systems,
which allow Agilent to provide this service at competitive
prices.
Calibration Interval
A 1- year interval is adequate for most applications. Accuracy
specifications are warranted only if adjustment is made at
regular calibration intervals. Accuracy specifications are not
NOTE
Make sure you have read “Test Considerations” on page 128 before
calibrating the instrument.