Features and Functions 3
Agilent U1251A/U1252A User’s and Service Guide 65
Decibel Display
The dBm operation calculates the power delivered to a
reference resistance relative to 1 mW, and can be applied to
DC V, AC V and AC + DC V measurements for decibel
conversion. Voltage measurement is converted to dBm by
using the following formula:
The reference resistance may be selected from 1~9999Ω in
Setup mode. Default value is 50Ω.
The decibel of voltage is calculated with respect to 1 V. The
formula is according to the voltage measurement below:
dBV = 20 log
10
Vin
1 At , or rotary switch position, push
to scroll to dBm measurement on the primary
display. AC voltage measurement is indicated on the
secondary display.
2 Push for more than 1 second to exit this mode.
NOTE
If rotary switch is at “~ V” position, push to switch between dBV
and dBm measurements. The dBm or dBV measurement can be selected
at ACV position, the selection will be the reference for other voltage
measurement.

Table of Contents

Other manuals for Agilent Technologies U1251A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1251A and is the answer not in the manual?

Agilent Technologies U1251A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1251A
CategoryMultimeter
LanguageEnglish

Related product manuals