Specifications 8
Agilent U1251A/U1252A User’s and Service Guide 161
U1252A AC Specifications
Ta b l e 7 U1252A AC Accuracy ± (% of reading + No. of Least Significant
Digit)
[1] The additional error to be added as frequency >20 kHz and signal input<10 %of
range: 3 counts of LSD per kHz.
[2] Input impedance: > 1GHz for 50 mV to 1000 mV. Input impedance is 1.1 MΩ
(nominal) with <100 pF for 5 V to 1000 V ranges.
[3] The input signal is lower than the product of 20,000,000 V-Hz (product of voltage
and frequency).
[4] Input current >35 μArms.
[5] Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5%
needs to be added to the specified accuracy if the signal measured is in the
range of 10 A to 20 A for 30 seconds maximum. After measuring a current of
>10 A, leave the meter to cool down for twice the measuring time used before
application of low current measurement.
[6] Input current < 3 Arms.
Function Range Resolution
Frequency
20 Hz
45 Hz
45 Hz
1 kHz
1 kHz
10 kHz
10 kHz
20 kHz
20 kHz
100 kHz
(1)
True RMS
AC Voltage
(2)
50.000 mV 0.001 mV 1.5+60 0.4+40 0.7+40 0.75+40 3.5+120
500.00 mV 0.01 mV 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
1000.0 mV 0.1 mV 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
5.0000 V 0.0001 V 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
50.000 V 0.001 V 1.5+60 0.4+25 0.4+25 0.75+40 3.5+120
500.00 V 0.01 V 1.5+60 0.4+25 0.4+25 1.5+40
3.5+120
(3)
1000.0 V 0.1 V 1.5+60 0.4+40 0.4+40
1.5+40
(3)
N/A
Function Range Resolution
Frequency
20 Hz 45 Hz 45 Hz 1 kHz 1 kHz 20 kHz
20 kHz
100 kHz
(1)
AC Current
500.00 μA
(4)
0.01 μA 1.0+20 0.7+20 0.75+20 5+80
5000.0 μA0.1 μA 1.0+20 0.7+20 0.75+20 5+80
50.000 mA 0.001 mA 1.0+20 0.7+20 0.75+20 5+80
440.00 mA 0.01 mA 1.0+20 0.7+20 1.5+20 5+80
5.0000 A 0.0001 A
1.5+20
(6)
0.7+20 3+60
<3 A/5 kHz
N/A
10.000 A
(5)
0.001 A
1.5+20
(6)
0.7+20

Table of Contents

Other manuals for Agilent Technologies U1251A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1251A and is the answer not in the manual?

Agilent Technologies U1251A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1251A
CategoryMultimeter
LanguageEnglish

Related product manuals