Performance Tests and Calibration 7
Agilent U1251A/U1252A User’s and Service Guide 127
Charge Terminal Alert Test
This test determines if the charge terminal alert is operating
properly.
The meter sounds an alert when the terminal
detects a voltage level of more than 5V but rotary switch is
not set to position. The meter sounds an alert beep
and primary display indicates and flashes “Ch.Err” until the
lead is removed from terminal.
Figure 56 Charge terminal alert
NOTE
Before conducting this test, make sure the beep function is not disabled in
setup.