132 Agilent U1251A/U1252A User’s and Service Guide
7 Performance Tests and Calibration
Step Test Function Range 5520A Output
Error from nominal 1 year
U1251A U1252A
4 Turn the rotary switch to
V position (for model U1252A), to
V position (for model
U1251A)
5 V 5 V ± 2 mV ±1.75 mV
50 V 50 V ± 20 mV ± 17.5 mV
500 V 500 V ± 200 mV ± 200 mV
1000 V 1000 V ± 800 mV ± 800 mV
5 Press button to go to the
V mode
[1]
5 V 5 V,1 kHz N/A ± 22.5 mV
5 V, 10 kHz N/A ± 22.5 mV
5 V, 20 kHz N/A ± 41.5 mV
5 V, 100 kHz N/A ± 187 mV
50 V 50 V, 1 kHz N/A ± 225 mV
50 V, 10 kHz N/A ± 225 mV
50 V, 20 kHz N/A ± 415 mV
50 V, 100 kHz N/A ± 1.87 V
500 V 500 V, 1 kHz N/A ± 2.25 V
500 V, 10 kHz N/A ± 2.25 V
1000 V 1000 V, 1 kHz N/A ± 8.0 V