BPAP Series User Manual V1.0
7 / 58
Maximum Flow
Test Pressures (hPa)
4
10
15
20
Measured Pressure at the
Patient Connection Port (hPa)
3
9
14
19
Average Flow at the Patient
Connection Port (L/min)
90
150
150
150
SpO
2
Range: 35% 100%
The margin of error for SpO
2
between 70% and 100% is ± 3%. No strict accuracy
requirements for SpO
2
below 70%.
Pulse Rate
Range: 30 240 BPM
Margin of Error: ± 2%
Wavelengths
Red: 663 nanometers
Infrared: 890 nanometers
Maximal Optical Output Power
Less than 1.5 mw maximum average.
Air Tubing
Air tubing
Length
Inner diameter
Tubing
6 ft.(1.83m)
19mm
Heater Tubing
6 ft.(1.83m)
19mm
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.
PM2.5 Filter
Efficiency: 90% for 2.5 micron dust