BPAP Series User Manual V1.0
19 / 58
There are five heat levels available, and the number of heat level will appear in the Main
Interface on the screen of the device. The number 3 next to the icon indicating the heat
is adjusted to Level 3, as shown in Fig. 11-10.
Fig. 11-10
(3) Connect the other end of the tubing to the mask according to the user manual for the
mask. Wear the mask.
WARNINGS!
• If multiple persons are going to use the device (e.g., rental devices), a low-resistance, main
flow bacteria filter should be installed in-line between the device and tubing. Pressures must
be verified by your home care provider when alternate or optional accessories are in place.
• If you are using a mask with a built-in exhalation port, connect the mask’s connector to the
tubing.
• If you are using a mask with a separate exhalation port, connect the tubing to the
exhalation port. Position the exhalation port so that the vented air is blowing away from your
face. Connect the mask’s connector to the exhalation port.
• If you are using a full-face mask (a mask covering both your mouth and nose), the mask
must be equipped with a safety (entrainment) valve.
• In order to minimize the risk of CO
2
rebreathing, the patient should observe the following
instructions:
- Use the accompanying tubing and mask provided by BMC.
- Do not wear the mask for more than a few minutes while the device is not operating.
- Use only masks with vent holes. Do not block or try to seal the vent holes in the exhalation
port.
11.6 Using Oxygen with the Device
Oxygen may be added at the mask connection. Please observe the instructions listed below
when using oxygen with the device.
WARNINGS!
• Connect the oxygen tube to the oxygen inlet of the mask.
• The oxygen supply must comply with the local regulations for medical oxygen.