BPAP Series User Manual V1.0
39 / 58
17.5 Alarming Information and Description
Alarm Message
Alarm
Priority
Alarm
Type
Description
Power Failure!!!
High
Priority
Technology
Alarm
An audible alarm will sound in 6s if the
device is accidentally disconnected from
power when it is delivering air. Alarming
duration time is no less than 30s.
Note:
(1) The alarm will not sound if power
failure occurs when the device is in
standby state.
(2) No alarm message on the screen
during a power failure
Device fault!!!
High
Priority
Technology
Alarm
An audible alarm will sound if no airflow
comes out of the machine; the screen will
display Device fault!!!
Tube
disconnected!!!
(only applies to
G3 B25VT, G3
B30VT, G3 B30SV
and G3 LAB)
High
Priority
Function
Alarm
When the airflow is on, an audible alarm
will sound if the tube accidentally
detached, the screen will display Tube
disconnected!!!
High Pressure!!!
High
Priority
Function
Alarm
When the airflow is on, an audible alarm
will sound if the airway pressure exceeds
the warning threshold; the screen will
display High Pressure!!!.
Note: The thresholds for different
models:
Off, 5 26 hPa applies to G3 B25VT, in
0.5 hPa increments, the default setting is
25 hPa.
Off, 5 31 hPa applies to G3 B30VT, G3
B30SV and G3 LAB, in 0.5 hPa
increments, the default setting is 30
hPa

Table of Contents

Other manuals for BMC G3 LAB

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC G3 LAB and is the answer not in the manual?

BMC G3 LAB Specifications

General IconGeneral
BrandBMC
ModelG3 LAB
CategoryOxygen Equipment
LanguageEnglish

Related product manuals