17.5 Alarming Information and Description
An audible alarm will sound in 6s if the
device is accidentally disconnected from
power when it is delivering air. Alarming
duration time is no less than 30s.
Note:
(1) The alarm will not sound if power
failure occurs when the device is in
standby state.
(2) No alarm message on the screen
during a power failure
An audible alarm will sound if no airflow
comes out of the machine; the screen will
display ―Device fault!!!‖
Tube
disconnected!!!
(only applies to
G3 B25VT, G3
B30VT, G3 B30SV
and G3 LAB)
When the airflow is on, an audible alarm
will sound if the tube accidentally
detached, the screen will display ―Tube
disconnected!!!‖
When the airflow is on, an audible alarm
will sound if the airway pressure exceeds
the warning threshold; the screen will
display ―High Pressure!!!‖.
Note: The thresholds for different
models:
Off, 5 ~ 26 hPa applies to G3 B25VT, in
0.5 hPa increments, the default setting is
―25 hPa‖.
Off, 5 ~ 31 hPa applies to G3 B30VT, G3
B30SV and G3 LAB, in 0.5 hPa
increments, the default setting is ―30
hPa‖