BPAP Series User Manual V1.0
47 / 58
18.8 Disinfection
Generally speaking, if you have strictly followed the above cleaning instructions, you do not
have to disinfect the device and / or water chamber. If the device is contaminated or used in
clinical trials, you may purchase disinfectants from a home medical equipment company to
disinfect the device.
Disinfection of the Water Chamber:
Prior to disinfection, clean the water chamber according to Section 18.3 ―Cleaning the Water
Chamber‖. The disinfection methods are as follows:
(1) Heat disinfection: Disinfect the water chamber by immersing it in tap water at 75°C ±2°C
for 30 minutes.
(2) Use mild disinfectants.
Disinfection of the SpO
2
Probe:
For details, refer to the disinfection instructions in the user manual for the SpO
2
Kit.
CAUTIONS!
• Disinfectants tend to damage materials and reduce the life of components. Try to select the
appropriate disinfectant, and follow the disinfectant manufacturer's instructions and
recommendations.
• After disinfection, check the disinfected component for any signs of damage. Replace any
damaged component immediately.
WARNINGS!
• After disinfection, rinse any disinfected component in clean water thoroughly, especially
components in close contact with the patient such as the mask, headgear, and tubing, so as to
prevent disinfectant residuals from damaging the skin or respiratory tract or causing allergies.
• The device shall not be serviced or maintained while in use with a patient.
• Sterilization of this device and its components other than recommended is not permitted.