14
POWER-UP ZERO FEATURE
FILTERING
Press ENTER to show the current setting. Press 0 or 1 to
enter the new setting and press ENTER to save it or to
save the current setting.
no = No Re-Zero on Power-Up YES = Automatic Re-Zero on Power-Up
Press ENTER to show the current setting. Press 0
through 3 to enter the new filtering level. Then press
ENTER to save it or to save the current filter setting.
0 = No Filtering 2 = Moderate Filtering
1 = Minimal Filtering 3 = Custom Filtering
If you selected 0, 1 or 2, advance to
MOTION RANGE.
PUO=
FLt=
F YOU SELECTED LEVEL 3 FILTERING
THEN:
BREAK RANGE
MOTION RANGE
BLANK ON MOTION
SAMPLE RATE
Press ENTER to show the current filter setting. Press 0
through 9 to enter the new filter value (1 to 16). The larger
the number, the greater the effect. Then press ENTER to
save it or to retain the current setting.
Filter is the number of weigh samples that will be averaged before being
displayed. If F=6, the weight value placed in the display will be the average of the
last six (6) consecutive weigh samples.
Press ENTER to show the current break range setting.
Press 0 through 9 to enter the new break range (1 to 64).
The larger the number the greater the weight change
required to break out of the filtering. Press ENTER to save
the new value or to save the current displayed value of
break range.
Press ENTER to show the current setting. Press 1 through
9 to enter the new value for the motion range. Weight
changes greater than this number of divisions will indicate
motion. Press ENTER to save the new setting or to save
the current motion range setting.
Press ENTER to show the current setting. Press 0 or 1 to
enter the new setting. Then press ENTER to save it or to
save the current setting.
no = Do Not Blank Display On Motion YES = Blank Display On Motion
Press ENTER to show the current sample rate. Press 0
through 9 to enter the new Sample Rate (1 to 12 samples
per second). Then press ENTER to save it or to save the
current setting.
Press ENTER to show the number of presets. Press 0
through 9 to enter the new number of preset weight
comparators. Then press ENTER to save it or to save the
current setting. If the number of preset weight comparators
is set to none (0 entered), advance to the DIGITAL FILL
CONTROL setting, otherwise continue.
F=
br=
UnSt=
bL=
Sr=
PSEtS