26 Bottle Setup & Conguration
5. Enter the solvent parameters: Name, Max Temp, and Cooling Temp.
• The Max Temp is the maximum temperature a method using the solvent can be set to.
WARNING
Ensure that the pressure of the solvent at the maximum set temperature is less than 200 psi. If unsure of
the Max Temp and Cooling Temp parameters, please contact CEM Corporation.
• The Cooling Temp is the maximum temperature at which the system will start another run using the solvent.
6. Determine if the solvent is preferred, and toggle the switch to ON or OFF.
• Only one solvent can be the preferred solvent.
• When creating a method the preferred solvent will auto-ll but can be changed to another solvent.
• The preferred solvent will be green in the “Solvent” list.
7. Select “OK.”
8. Select the Save icon .
CAUTION
The EDGE operates in a sealed environment allowing it to perform at an elevated temperature and
pressure. The maximum temperature of the solvent is higher than the boiling point.
Edit Solvent
1. Select the System Menu icon in the bottom right corner of the screen.
2. Select Settings.
3. Select Solvents.
4. Select the solvent to edit.
5. Select the Edit icon at the top right of the screen.
6. Make desired changes to the “Name”, “Max Temp”, “Cooling Temp”, and whether or not the solvent is “pre-
ferred.”
7. Select “OK.”
8. Select the Save icon at the top right of the screen.