42 Calibrations
NOTE
Pay attention to the dispense needle during this time; it should enter into the waste port smoothly. If it
doesn’t, the waste position will need to be recalibrated.
6.1. Select “Home.” The autosampler will travel to the home position.
6.2. Scroll upward to the Actuator section.
6.3. Select “Open.”
NOTE
If the autosampler crashes or movement is not smooth at any point during the verication, the
autosampler will need to be calibrated. Please see “Autosampler Calibration.”
Pressure Calibration
The pressure requires calibration during initial install, if the EDGE system is moved, and semi-annually.
1. Select the System Menu icon in the bottom right corner of the screen.
2. Select Tools.
3. Select Calibration.