EasyManuals Logo
Home>CEM>Laboratory Equipment>EDGE system

CEM EDGE system User Manual

CEM EDGE system
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
60 System Specications
System Specifications
Location Requirements
• Has adequate ventilation that can be accessed via the exhaust hose (no further than 10’ away) to remove fumes
from system
• Provides adequate space for sample handling
• Has a dedicated, grounded outlet no more than 5 ft (1.5 m) from unit
• Is free from vibration of large equipment
• Provides clear visual access to the display of the system
• The system weighs 70 lbs (31.75 kg) without rack or vessels
• Instrument Dimensions: 14.25” (37 cm) W x 25.53” (65 cm) D x 27.93” (71 cm) H
• Recommended counter space: 45” (115 cm) W x 30” (77 cm) D x 28” (71 cm) H
• The above provides 12” (31 cm) of open space on the right side of the system so that the racks can be
easily loaded and 18.75” (48 cm) on the left side of the system for sample preparation and access to the
power switch
• Do not position the EDGE so it is difcult to access the power switch
Electrical Requirements
• Region Specic
• 100 - 120 VAC, 50/60 Hz, 10 A
or
• 200 - 240 VAC, 50/60 Hz, 5 A
• Line voltage must not vary more than 10% of its specied level
• Dedicated, grounded outlet
• Instrument is supplied with a power cord or an adequately rated power cord may be used
• Fuses:
• 100 - 120 VAC: 250 V, 10A 5 mm x 20 mm, fast acting
or
• 200 - 240 VAC: 250V, 5 A, 5 mm x 20 mm, fast acting
Environmental Requirements
• Indoor use only
• Altitude up to 2000 m
• Operating Temperature range of 50°F (10°C) to 85°F (29.4°C)
• A fume hood within 10 feet of unit or adequate exhaust that pulls a minimum of 30.5 CFM at point of connection
• Instrument is supplied with a ten foot long exhaust hose with a 2” ID
• Relative humidity range of 10 % to 85%, non-condensing
• Overvoltage Category II
• Pollution degree 2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM EDGE system and is the answer not in the manual?

CEM EDGE system Specifications

General IconGeneral
BrandCEM
ModelEDGE system
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals