53System Repacking
System Repacking
Prepare to Ship
Follow procedures below to prepare the EDGE for shipment. If shipping brackets are not available, please contact
CEM Corporation.
1. Remove the rack from the EDGE platform.
CAUTION
Never ship unit with rack in position.
2. Select the System Menu icon located in the lower right corner of the screen.
3. Select Tools.
4. Select Utilities.
5. Select Prepare to Ship “Start.
6. Follow prompts on the software screen to prime lines with water, ush chamber, dry lines, and add shipping
brackets. If shipping brackets are not available, please contact CEM Corporation.
CAUTION
Solvent lines must be purged prior to packing or shipping instrument.
7. Select “OK.
8. Follow the directions on the screen to align with the shipping brackets.
NOTE
Screws should be installed on the instrument already. Shipping brackets should be with the shipping boxes
the system was delivered in. Remove appropriate screws, place bracket in position and re-install screws.
9. Once shipping brackets are installed, turn the EDGE off using the power switch located on the left side of the
unit.
10. Disconnect the power cord from the EDGE and power receptacle.
11. Ensure all bottles and waste lines have been removed.