104
The instrument displays the weight of the
sample, the weight of the sample and diluent,
and the dilution ratio.
30. Press READY to continue the analysis.
31. Lift the cover of the SMART System
5
.
Place one round and two square glass fiber
sample pads on the balance pan (round pad
on bottom). Close the instrument cover.
32. Press TARE.
33. Wait for the instrument to tare the weight
of the sample pads.
Note: The “T” in the lower left corner of the
balance icon indicates that the weight of the
sample pads has been tared.
34. Lift the instrument cover. Remove the
square sample pads from the balance pan.
35. Apply the diluted sample in a thin, even
layer to one of the square sample pads and
cover the sample with the other square
sample pad.
36. With the round pad still on the balance pan,
quickly and gently place the sample pads
back on the balance pan. Close the
instrument cover.
Dilution Moisture/Fat (Ext)
Sample Wt: xx.xxxg
Sample + Dil Wt.: xx.xxxg
Dilution Ratio: x.xxxx
Press READY to continue.
XXXXXX
Power: xxx %
Temp (xxx): xx.x
Time: xx:xx
Ratio: x.xxxx
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID RATIO DILUTION
1 2 3 4 5
TARING BALANCE . . .
XXXXXX
Power: xxx %
Temp (xxx): xx.x
Time: xx:xx
Ratio: x.xxxx
Wt: x.xxxxg
Place sample and pads on
balance and press START.
ID RATIO DILUTION REDRY
1 2 3 4 5
T

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM SMART System 5 and is the answer not in the manual?

CEM SMART System 5 Specifications

General IconGeneral
BrandCEM
ModelSMART System 5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals