Current Loop Connection
Sensors using current loops are widely used in industry to communicate analog signals in the presence
of electromagnetic interference. Both 2- and 3-wire current loops (often referred to as loop powered or
separately excited, respectively) are commonly used in industry. Both types of current sensors are
illustrated in this section. Internally, the ELITEpro XC uses 0.1% precision 499 ohm resistors to measure
the voltage drop impressed by the external current source. Current loop sensors will typically be
powered from a 24-volt DC supply. The ELITEpro XC has four measurement channels. The negative
terminals of each channel are common to each other and connected to the reference plane for power
measurements. For this reason it is imperative that the ELITEpro XC be connected as the last component
in the current loop rather than the first if multiple channels are used. Best practices are to use a single
power supply for all sensors to reduce the occurrence of ground loop current between supplies.
L1
L2
G
+
-
Sensor
Power Supply
2-W
ire Current Loop Configuration
w ww. . com
information@itm.com1.800.561.8187

Table of Contents

Other manuals for DENT Instruments ELITEpro XC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DENT Instruments ELITEpro XC and is the answer not in the manual?

DENT Instruments ELITEpro XC Specifications

General IconGeneral
BrandDENT Instruments
ModelELITEpro XC
CategoryData Loggers
LanguageEnglish