Appendix
13 Information about EMC in accordance with EN 60601-1-2
13.1 General notes
The information in this leaflet includes excerpts from the relevant European standards for electrical,
medical devices. It must be observed when installing Dürr Dental devices or combining them with prod-
ucts of other manufacturers. If you are uncertain about anything, please refer to the complete standard.
13.2 Abbreviations
EMC
Electromagnetic compatibility
HF High frequency
U
T
Rated voltage of the device (supply voltage)
V
1
, V
2
Compliance level for the test in acc. with IEC 61000-4-6
E
1
Compliance level for the test in acc. with IEC61000-4-3
P Rated power of the transmitter in Watts (W) in accordance with the specifications of the
transmitter manufacturer
d Recommended safety distance in metres (m)
13.3 Guidelines and manufacturer's information
Electromagnetic emissions for all devices and systems
The device is designed for operation in an electromagnetic environment as specified below. The cus-
tomer or operator of the device should ensure that the device is operated in such an environment.
Interference emission
measurements
Compliance Electromagnetic environment - guidelines
HF emissions in accord-
ance with CISPR 11
Group 1 The unit uses HF energy exclusively for internal functions.
As a result, HF-transmissions are very low and it is highly
unlikely that any interference will be caused to neighbour-
ing electronic devices.
HF emissions in accord-
ance with CISPR 11
Class A The VistaIntra DC is suitable for use in facilities other than
living areas and those areas that are directly connected to
the PUBLIC MAINS ELECTRICITY SUPPLY that also sup-
plies buildings used for residential purposes.
Harmonics in acc. with
IEC 61000-3-2
Not applica-
ble
Voltage fluctuations/flickers
in acc. with IEC 61000-3-3
Not applica-
ble
Appendix
2202100028L02 2007V019 31
EN