13.4 Calculation table
If the measured values deviate from the standard, the values are specified in chapter "Electromagnetic
emissions for all devices and systems".
The safety distances can then be calculated in the tables shown below.
P: ..........
V
1
: ..........
E
1
: ..........
P Rated power of the transmitter in Watts (W) in accordance with the specifications of the
transmitter manufacturer
V
1
Compliance level for the test in acc. with IEC61000-4-6
E
1
Compliance level for the test in acc. with IEC61000-4-3
Interference
immunity tests
IEC 60601 - test
level
Compliance level Recommended safety distan-
ces
Conducted HF dis-
turbance variables in
accordance with
IEC 61000-4-6
3 V
eff
150 kHz to 80 MHz
[V
1
] V d = [3.5 / V
1
] × ÖP
Emitted HF disturb-
ance variables in
accordance with
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
[E
1
] V/m d = [3.5 / E
1
] × ÖP
for 80 MHz to 800 MHz
d = [7 / E
1
] × ÖP
for 800 MHz to 2.5 GHz
Rated power of the
transmitter (W)
Safety distance based on the transmission frequency (m)
150 kHz to 80 MHz
d = [3.5/V
1
] ×ÖP
80 MHz to 800 MHz
d = [3.5/E
1
×ÖP
800 MHz to 2.5 GHz
d = [7 / E
1
] ×ÖP
0.01
0.1
1
10
100
Appendix
2202100028L02 2007V019 35
EN