SD3 Series Ultrasonic Pocket Doppler User Manual Troubleshooting
- 44 -
A4.6 Transducer Acoustic Output Parameters List
Acoustic output reporting table for IEC60601-2-37
(IEC60601-2-37, Edition 2.1, 2015-06, table 201.103)
Transducer Model: SD3 CW2.0
Working Mode: CW mode
Working Frequency:2.0MHz
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.019
0.16
0.40
N/A
Index component value
N/A
0.16
N/A
0.40
Acoustic
Parameters
p
r.
α
at z
MI
(MPa)
0.027
P (mW)
24.21
24.21
N/A
P
1x1
(mW)
N/A
N/A
z
s
(cm)
2.30
z
b
(cm)
2.73
z
MI
(cm)
2.80
z
PII.
α
(cm)
2.80
f
awf
(MHz)
2.00
2.00
2.00
N/A
Other
Information
prr (Hz)
N/A
srr (Hz)
N/A
n
pps
N/A
I
pa.
α
at z
PII.
α
(W/cm
2
)
0.026
I
spta.
α
at z
PII.
α
or z
SII.
α
(mW/cm
2
)
26.49
I
spta
at z
PII
or z
SII
(mW/cm
2
)
42.63
p
r.
at z
PII
(MPa)
0.037
Operating
control
conditions
Focus(mm)
Fixed
Depth(mm)
Fixed
Frequency(MHz)
2.00