Appendix 5 Troubleshooting
The speaker does not sound.
Bad connection between the
probe and main unit.
Make sure that the probe is well
connected to the main unit.
No coupling gel is applied.
Apply coupling gel on the probe
acoustic surface.
The battery is in low energy.
Replace the battery or charge the
rechargeable battery.
The probe is not placed at
the optimum position.
Locate the probe at a position
for optimal FHR examining.
No coupling gel is applied.
Apply coupling gel on the probe
acoustic surface.
The screen displays nothing
and the speaker does not
sound after the main unit is
switched on.
The battery is in low energy.
Replace the battery or charge the
rechargeable battery.
The speaker sounds but the
screen displays nothing or
displays wrong characters
after the main unit is
switched on.
The device is subject to
strong electromagnetic
emissions.
Make sure that the environment
in which the device is operated
is not subject to any source of
strong electromagnetic
emissions.
The probe has not been
taken out from the probe
holder when the main unit is
switched on.
Take out the probe from the
probe holder before switching
on the main unit.
The battery is in low energy.
Replace the battery or charge the
rechargeable battery.
Too much coupling gel on
the probe acoustic surface.
Wipe the residual coupling gel
off the probe after each use.
The screen displays nothing
Replace the power adapter.