SD3 Series Ultrasonic Pocket Doppler User Manual Troubleshooting
- 49 -
A4.7 Track1 Acoustic Output Range Summary
TRACK1 SUMMARY
Mode of Operation
Transducer
Global Maximum Output Level
CW mode
CW2.0
Max I
SPTA.3
(mW/cm2)
35.4722
Min I
SPTA.3
(mW/cm2)
17.5000
Max I
SPPA.3
(W/cm2)
0.03547
Min I
SPPA.3
(W/cm2)
0.01750
Max MI
0.0225
Min MI
0.0162
Max I
sata@
transducer face
12.77
Min I
sata@
transducer face
5.27
CW3.0
Max I
SPTA.3
(mW/cm2)
30.8334
Min I
SPTA.3
(mW/cm2)
19.2886
Max I
SPPA.3
(W/cm2)
0.03083
Min I
SPPA.3
(W/cm2)
0.01929
Max MI
0.0210
Min MI
0.0146
Max I
sata@
transducer face
19.91
Min I
sata@
transducer face
15.38
CW4.0
Max I
SPTA.3
(mW/cm2)
18.8302
Min I
SPTA.3
(mW/cm2)
14.8597
Max I
SPPA.3
(W/cm2)
0.01883
Min I
SPPA.3
(W/cm2)
0.01486
Max MI
0.0125
Min MI
0.0112
CW5.0
Max I
SPTA.3
(mW/cm2)
32.3369
Min I
SPTA.3
(mW/cm2)
26.7624
Max I
SPPA.3
(W/cm2)
0.03234
Min I
SPPA.3
(W/cm2)
0.02676
Max MI
0.0139
Min MI
0.0117
CW8.0
Max I
SPTA.3
(mW/cm2)
10.7050
Min I
SPTA.3
(mW/cm2)
8.2472
Max I
SPPA.3
(W/cm2)
0.01071
Min I
SPPA.3
(W/cm2)
0.008247
Max MI
0.0059
Min MI
0.0053

Table of Contents